Signal flow overview
Figure 5-4 LOGIQ E9 Signal Flow
GRLY
R
e
l
a
y
s
The MRX Board combines the functionality of the Receiver Boards (DRX and GRX) and the GFI.
The Transmit Pulse bursts transmitted, are still routed from the GTX board(s) via the XD bus to the
Relay board where the ultrasound probes send the energy into the body. Weak ultrasound echoes from
body structures and blood cells are received by the probes and routed via the Relay board and the XD
bus to the MRX board. The MRX board amplifies the ultrasound signal and connects it with an A/D
converter to the digital domain. The digital signals are then further processed on the MRX boards.
All the processing is the same as a GFI configuration, but now the DRX, GRX and GFI are no longer
separate boards.
5 - 6
. 2
EV
Frontplane
MRX
MRX 192 Ch
TGC
TGC
TGC
256
TGC
T/ R
TGC
Section 5-2 - LOGIQ E9 description
- MRX
256
CW
Option
A/D
Converters
I / Q
Filter
X
I
Analog
Beamformer
X
Q
X
A/DA
Nathan
A/DA
32 Ch
Nathan
A/DA
32 Ch
A/DA
Nathan
32 Ch
Nathan
32 Ch
Nathan
32 Ch
Nathan
32 Ch
A/D
LOGIQ E9 S
ERVICE
GTX 64 Ch
GTX 64 Ch
GTX
David
Pulser
64 Ch
0
David
1
David
2
David
Pulser
3
GDIF
Front End
Interface
DSP
SED
TGC
Test
GFE
Gen
Gen
M