GE LOGIQ E9 Service Manual page 269

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5573152-100, R
IRECTION
7-5-9-6
Video Card
See:
7-5-3 "Diagnostics Window Overview" on page 7-16
Runs a test on essential functions of the Video Card. The test results are displayed in the Status portion
of the window.
This diagnostic tests your system's video capabilities. This involves testing the video memory with 18
patterns, testing your graphics acceleration, and text output. You will see these tests being performed
on your monitor.
You can cancel this test at any time by hitting the Escape (Esc) key.
The following tests are performed on your monitor:
Memory - Video memory is tested by filling the video buffer with 18 test patterns, one pattern at a
time. The tests will fill the entire screen with a single color.
Data Transfer - This tests the graphics acceleration part of your video controller. These tests will
appear on your screen as black and white concentric squares and rectangles of various sizes and
colors. If errors are detected, the locations of the problems are displayed.
Text Output - This test prints a text string in random sizes and colors to tests your video device driver
and video controller.
Run Time: May take about 10 minutes.
1.) Enter the desired Loop Count (numeric value of 4 digits or less).
2.) Click the Execute button.
3.) Replace the Video Card if there are any errors. If the problem continues:
4.) Replace the BEP.
7-5-9-7
S-Video Image displays a portion of the Main Display
The S-Video image displays a portion of the main display that includes the probe image in both single
and dual probe display modes. To provide increased resolution and maintain the proper aspect ratio on
the S-Video display, the entire main display image is not projected on the S-Video monitor. The S-Video
crop area was selected to optimize the probe image so when the LOGIQ E9 is not in an imaging mode,
the S-Video will continue to display a cropped portion of the screen which may appear incorrect.
7-5-9-8
The Monitor Measurement Summary Sidebar does not appear after Upgrading to R6
After upgrading to R6, the monitor Measurement Summary Sidebar does not appear, and the image
uses the same space it would for Dual, all the time. So the image will go right up to the left edge of the
screen.
When upgrading to R6 software in a LOGIQ E9 with a 19 inch LCD, the following steps must be
completed to have the correct layout for 19 inch Monitors. DO NOT skip this step, or the monitor layout
will not be correct.
Go to Utilities -> System -> System Display:
-
-
. 2
EV
Set the drop-down menu in Use Wide Screen for... > Single Image to OFF
Make sure the checkbox in Side Panel Content > Measurement Summary is checked.
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 S
for a description of this window's features.
M
ERVICE
ANUAL
7 - 67