GE LOGIQ E9 Service Manual page 31

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

GE
D
5573152-100, R
IRECTION
1-4-2
Human safety (cont'd)
WARNING
WARNING
When the top console is in its locked position, the gas shock is compressed and
stores mechanical energy. During normal operation the top console, the weight of
the monitor and the mechanical force of the gas shock are in balance. Take care if/
when you activate this gas shock. Personal injury can occur after the panel is
removed and the shock pressure is released. Take care when you repair the
elevation assembly.
Ensure that the system is turned off and unplugged.
WARNING
WARNING
Wait for at least 20 seconds for capacitors to discharge as there are no test points to verify isolation.
The Amber light on the Op Panel On/Off button will turn off.
Beware that the Main Power Supply, Extended Power Shutdown and Back End Processor may be
energized even if the power is turned off when the cord is still plugged into the AC Outlet
Risk of electrical shock, system must be turned off and disconnected from power
WARNING
WARNING
source. Cord must be controlled at all times.
discharge as there are no test points to verify isolation. The Amber light on the Op Panel On/Off
button will turn off.
Beware that the Main Power Supply, Extended Power Shutdown and Back End Processor may be
energized even if the power is turned off when the cord is still plugged into the AC Outlet
Use extreme caution as long as THE LOGIQ E9 is un-stable, not resting on all four
WARNING
WARNING
Casters.
WARNING
WARNING
Tilting the console requires two people in order to avoid injury to service personnel
and damage to the equipment.
WARNING
WARNING
Beware of possible sharp edges on all mechanical parts. If sharp edges are
encountered, the appropriate PPE should be used to reduce the risk of injury.
The appropriate PPE is required per EHS Policies and SRA's.
WARNING
WARNING
Wear all PPE including gloves as indicated in the chemical MSDS.
. 2
EV
Chapter 1 Introduction
LOGIQ E9 S
Wait for at least 20 seconds for capacitors to
M
ERVICE
ANUAL
1 - 17

Advertisement

Table of Contents
loading

Table of Contents