8-5-4
Shearwave Console Cover replacement
NOTE:
Whenever the left Side Cover is removed, clear the BEP Fan(s) intake(s) of any debris.
Manpower /
Time
One person /
10 minutes
Table 8-8
1. Power down the system.
2. Disconnect the mains power cable from the wall outlet and all Probes and External I/O Cabling.
3. Remove the Left Side Cover and loosen the Top Cover.
Preparation Links (if you need more information):
• 4-2-3 "Power shut down" on page 4-6.
• 8-5-3 "Side Covers replacement" on page 8-28.
• 8-5-5 "Top Cover replacement" on page 8-35.
. 2
EV
Table 8-7
Manpower / Time and Tools
Refer to:
8-2-5 "Tools needed for servicing the LOGIQ E9" on page
Preparations and Preparation Links
Preparations - you must perform the following steps
Chapter 8 Replacement procedures
LOGIQ E9 S
Tools
M
ERVICE
ANUAL
8-5.
8 - 33