1
Fig. 9
To prevent the other wheel from turning, secure it with several wedges.
Do not apply the parking brake.
Remove the wheel, four screws as shown in (Fig. 8).
Remove the six screws that hold the motor. Two of the screws can be accessed through
a hole in the toothed wheel.
Remove the complete motor unit inward through the opening in the chassis.
Install the new motor with the six screws.
Install the motor power unit (D102) or battery charger (D201). Refer to (Fig. 7) for
electrical connections.
Install the wheel with the four screws.
Remove the block of wood.
Reattach the covers.

Repairs to the motor power unit

The motor power unit should not be repaired. If the motor power unit is defective (with the
exception of 4 A fuse), the entire unit should be replaced.

Exchange of motor power unit D102

Remove the front cover.
Disconnect K101, 102, 104, 106 on D102.
Remove the two front screws (1/Fig. 9).
MOBILETT Plus HP
1
SPR8-220.061.03
Rev. 03
Motor drive units
Page 6 of 8
09.03
CS PS 24
11
11
Siemens AG