Battery circuit and power conversion unit
Fig. 9
Exchange of fuse U11
•
Pull out capacitor bank
(see section: "Pull out capacitor bank, M11" on Page 7 - 3).
•
Make sure the capacitor units (both M10 and M11) are discharged.
•
Replace U11 (pos. 1 / Fig. 9).
Exchange of capacitor bank M11
•
Remove power conversion unit
(see section: "Pull out power conversion unit, M10" on Page 7 - 1).
•
Remove capacitor bank
(see section: "Pull out capacitor bank, M11" on Page 7 - 3).
Exchange of capacitors of M11
•
Remove power conversion unit
(see section:"Pull out power conversion unit, M10" on Page 7 - 1).
•
Remove capacitor bank
(see section:"Pull out capacitor bank, M11" on Page 7 - 3).
1
MOB00627
Fig. 10
•
Remove the four screws (pos. 1 / Fig. 10).
•
Exchange the capacitors C1 and C2.
Siemens AG
Medical Solutions
+
+
SPR8-220.061.03
Rev. 03
1
1
Page 5 of 8
09.03
CS PS 24
7 - 5
MOB00632
7
7
7