Battery circuit and power conversion unit
DANGER
NOTE
Pull out power conversion unit, M10
Fig. 1
Remove left cover.
Remove the cover with the warning sign (pos. 1 / Fig. 1).
Disconnect K301/ K4A/ K4B/ K2 on D301.
Remove cable tie (pos. 1 / Fig. 2).
Siemens AG
Medical Solutions
High Voltage!
Life-threatening hazard of electric shock. Never work with the
system open if the batteries are still connected and the capacitor
is charged. If the batteries are connected, the complete system is
powered on!
The capacitors in M10 and M11 may still be charged even if the
system is switched OFF and the mains cable is
disconnected. The capacitors in M10 and M11 must be consid-
ered charged until the protective measures listed in this section
have been performed.
The power conversion unit is also called power transmission unit
in some documents.
2
1
1
SPR8-220.061.03
Rev. 03
09.03
Fig. 2
Page 1 of 8