9 - 4
Fig. 7
Replacing the collimator
Equipment required: rope
•
Switch off the unit and disconnect the power plug.
•
Turn the single tank so that the collimator is facing upwards (Fig. 7).
•
Secure the arm and the single tank with a rope (Fig. 7).
•
Remove covers "A" and "B".
•
Remove collimator cover "C".
MOBILETT Plus HP
Replacement and repairs on the collimator
C
A
B
SPR8-220.061.03
Rev. 03
Page 4 of 10
09.03
CS PS 24
9
Siemens AG
Medical Solutions