Siemens MOBILETT Plus HP Service Instructions Manual page 92

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

10 - 8
Remove the upper column cover by removing the transport handle and the lower column
cover by lifting it upwards.
Raise the single tank to maximum height and secure it with a ceiling support (Fig. 10).
CAUTION
NOTE
Remove the supporting wheels.
Put the protective mat under the unit to protect the floor from damage.
Shaft
Fig. 11
Remove battery charger D201 and motor power unit D102 (see section 7 and 11).
Disconnect all cables, ground connections and K68 according to (Fig. 11).
Remove the locking rings.
Remove the bearing rings.
Remove the shaft.
Let the turn plate down and remove it.
MOBILETT Plus HP
The ceiling suspension must be able to hold a weight of 100 kg
since the entire weight of the arm will be on the ceiling support
once the spring is removed.
If not, the arm and single tank will fall.
Make sure the ceiling support is sufficient.
Do not stand under the single tank or the arm system during the
work that follows.
SPR8-220.061.03
Rev. 03
09.03
Mechanical systems
K68
Page 8 of 28
CS PS 24
Medical Solutions
Siemens AG