Service programs

P07 - Query filament values from the battery-backed RAM

The kV values are displayed in the "kV" field. The corresponding filament values are dis-
played in the "mAs" field. Pressing the "kV +" button will display the next higher kV and
corresponding filament value and vice verse with "kV -".
The program starts at 40 kV in the table containing the filament values. The program
shows only the selected number of kV steps.
The kV is changed in steps by briefly depressing the "kV + or kV -" buttons. By keeping
the buttons pressed, the kV value will be continuously stepped up or down.
The generator filament values are self-adjusted to produce a kV value with minimum devi-
ation. If the software has been optimized, a filament value for a specific kV value will have
been indicated with the letter "A" in the extreme right position on the mAs display.

P08 - Display the message history

The history contains two parts. The first is the appearance number and the second the
actual message. The appearance number will be displayed in the "kV" field.
The corresponding message will be displayed in the "mAs" field. Pressing the "kV +" but-
ton will display the next higher number and corresponding message and vice verse with
"kV -".
The program starts by showing the latest message in the history. If there are no messages
the display will show:
kV mAs
0
The latest message has the highest number, for example:
kV mAs
20 CAL13 = There are total 20 new messages, number 20 is "CAL 13".
To exit P08, use the "lamp" button.
Siemens AG
Medical Solutions
Error = Message history empty
SPR8-220.061.03
Rev. 03
Page 5 of 8
09.03