10 - 2
Fig. 2
WARNING
MOBILETT Plus HP
MOB00484
Danger of personal injury !
Make sure the visible part of the thread between the link and the
key grip at the adjustment rod is always less than 35 mm (Fig. 3).
If the visible part of the adjustment rod is longer than specified,
the single tank and the arm might fall.
If any adjustment is needed, the single tank should be secured in
its uppermost position with a ceiling support (Fig. 2). After the
single tank has been secured, the two nuts can be loosened and
the adjustment rod can be turned.
Lock the rod with the nuts and repeat the parallelism check until
it is within the specified limit.
SPR8-220.061.03
Rev. 03
09.03
Mechanical systems
Fig. 3
Page 2 of 28