Replacement and repairs on the collimator

Light field relative to the X-ray field

Equipment required: Cassette, X-ray film and coins
NOTE
Load a 35 x 35 cm (14" x 14") cassette with film.
Place the cassette on a table or a similar surface.
Set a SID of 100 cm (39"). Use a measuring tape.
Switch on the collimator lamp and align the cassette. Collimate the light field to
25 x 25 cm (10" x 10").
Mark the four corners with coins and mark one of the corners with two coins (Fig. 5).
Set 52 kV, 2.0 mAs on the control panel.
CAUTION
Release an exposure.
Develop the film and measure the deviation between the light and the radiation field.
Acceptable deviation ± 1.7% of SID, (Fig. 6).
If the deviation is more than 1.7%, adjust with the screws marked C and D (Fig. 2) to
move the light field closer to the radiation field.
Repeat the test with a new exposure.
If necessary, adjust again until the deviation between the light field and the radiation field
is within the tolerance specified.
Reattach cover 1 (Fig. 1).
Siemens AG
Medical Solutions
If the MOBILETT is equipped with a DAP measuring system,
remove the DAP ionization chamber first.
X-ray radiation!
If not observed, risk of radiation exposure.
Use radiation protection!
SPR8-220.061.03
Rev. 03
09.03
Page 3 of 10