Display messages
Dis-
played
messages
CAL 7
Err 9
* See section "Monitoring the kV and mA values"
** Wiring diagram X037E
Tab. 1
Siemens AG
Medical Solutions
Mode of
Cause
operation
During
The exposure time
exposure
has reached its max-
imum value.
During
When measuring the
exposure
kV value, a large
high tension asym-
metry is detected.
SPR8-220.061.03
Rev. 03
Possible reasons and advice
The maximum exposure time is 4 s.
The exposure is normally
interrupted earlier by "Err 5".
Check the capacitance in the
capacitor bank (see section "P03 -
Capacitor bank and charging test").
Check the filament heating (see
section "P01 - Filament test").
Check the tube current in
measuring point "MAPOS" on D1
(see pages 140 and 141).
Check the current to the single tank
(see section "Error diagnosis HT
and single tank").
Normally the kV value is symmetric
around ground. The acceptable devi-
ation is ± 19.5 kV. This corresponds
to HTD 2.5 V± 0.5 V.
Intermittent measurement cables
or defective single tank.
Check the measurement cable K9
between D1 and the single tank
(see page 140**).
Check the current to the single tank
(see section "Error diagnosis HT
and single tank").
Measure the HTD signal on D1 and
compare it with notes and curves
on page 140 and 141**.