s27* PREP.
Rotating anode runs
Yes
Anode speed
reached
Yes
s27* EXP.
Yes
High voltage
"Exposure"
Yes
Exposure
interrupted
No
Correct
density
Yes
MOBILETT Plus HP
* S27 is the exposure release switch. First step = PREP, second step = EXP.
MOBILETT Plus HP
No
No
No
No
Yes
No
OK
SPR8-220.061.03
Rev. 03
Troubleshooting chart
Exposure release switch,
cables
and connections
Err 6 (see P02 - Rotation test)
Remote control (option)
USE 39
Err 6
Exposure release switch,
cables and connections
Displayed message...
Remote control (option)
Displayed message...
Displayed message...
kV-mAs-selection
Exposure time
Heating value
Darkroom
Page 2 of 2
09.03
CS PS 24
Err 6
Displayed
message
Displayed
message
Siemens AG