Appendix C:
Output may change depends on heat of LED and fluctuation of ambient temperature.
If CIS's output level change after white/dark correction, operate those correction again.
See reference data C.
*Regarding to the definition of the temperature of case : Tw, see item 2-7.
Reference data C : Correlation between CIS's output on continuous operation
and the temperature of case, Tw
100%
98%
96%
94%
92%
90%
0:00
TM-XG537C
CIS's output
Temp. of case
1:00
2:00
Continuous scan time 〔Hr〕
COPYRIGHT © 2020 MITSUBISHI ELECTRIC CORPORATION. ALL RIGHTS RESERVED.
Output on continuous operation
50
45
40
35
Tw
30
25
3:00
<Condition>
・ Ta=25 ± 5 ℃
・ LED-Duty = 65%
・ White-basis-chart
( O.D.=0.05 ~ 0.09)