Space Requirements; Scanning Room - Toshiba Aquilion TSX-101A/I Series Site Planning Manual

Table of Contents

Advertisement

5.

SPACE REQUIREMENTS

The floor space for unit installation should be 25 m
5.3 provides an example of the layout for the minimum space and subsection 5.4 provides a
reference figure for the maintenance space.
For patient couch CBTB-019A : Required space
For patient couch CBTB-019B : Required space
Patient couch model
CBTB-019A
CBTB-019B
5.1

Scanning Room

The gantry and patient couch are installed in the scanning room.
(1)
Installation of the gantry
The following space is required for installation and maintenance of the gantry and
inward/outward movements of the couch top:
A
(2)
X-ray shielding
Since the gantry releases X-rays, X-ray shielding should be provided around the scanner
room and the entrance.
(Refer to subsection 7.2 "X-ray Shielding in the Scanning Room".)
(3)
Ceiling height
The ceiling should be 2500 mm or higher at a floor level considering that equipment such as
contrast medium injection tools are installed on the ceiling.
Scanner room
5.60 m × 3.80 m
5.00 m × 3.80 m
3830 or more
C
(Rear)
(Gantry)
2330
(Front)
(Top view)
2
or more, divided into two rooms. Subsection
2
27 m
or more
2
25 m
or more
Operation room
1.50 m × 3.80 m
B
A:
B:
C: 1470 mm or more
13
No. 2A201-519EN*A
Minimum required
space
2
27 m
2
25 m
800 mm or more
700 mm or more
(For patient couch CBTB-019A)
1170 mm or more
(For patient couch CBTB-019B)