Scope; Application; System Composition - Toshiba Aquilion TSX-101A/I Series Site Planning Manual

Table of Contents

Advertisement

1.

SCOPE

This document applies to the Toshiba scanner Aquilion model TSX-101A.
Model
TSX-101A/I series
TSX-101A/H series
*1
: The patient couch used in combination depends on the system model.
2.

APPLICATION

This document describes site planning requirements to ensure optimal performance. It also
describes necessary site facilities and outlines the range of responsibilities during installation work.
3.

SYSTEM COMPOSITION

(1)
Standard composition
Gantry (X-ray high-voltage generator mounted type)
Patient couch
Console (CPU chassis (CPU BOX, REC BOX), monitors×2, keyboards×2, mice×2)
Accessories
Operation manual
(2)
Miscellaneous
Desks, such as for the console and patient observation system, and chairs are not included in
the system configuration.
The desks and chairs selected should meet the requirements below.
(a)
Desk
Height
Width
Depth
(b)
Chair
Seat height : The range from 500 to 550 mm should be covered.
High stability (5 legs)
With footrest or foot bar
Type
32-slice system
64-slice system
: 800 mm (When the CPU BOX is installed under the desk, a height of more
than 700 mm is required under the desk.)
: 1600 mm
: 800 mm
Standard patient couch
CBTB-019A or CBTB-019B
CBTB-019A or CBTB-019B
8
No. 2A201-519EN*A
*1
1 set
1 set
1 set
1 set
1 set