Precautions - Toshiba Aquilion TSX-101A/I Series Site Planning Manual

Table of Contents

Advertisement

7.5.3

Precautions

(1)
The pit and duct must be electrically shielded.
(2)
The pit and duct must be protected against flooding.
(3)
The cover should be divided into sections approximately 1 meter in length.
Floor surface
Iron cover
Iron sheet
(Fixed to the floor
using bolts)
100
100
Figure 7.5-1 Sectional view of a sample wiring pit
31
Divided into approximately
1 meter sections
Iron sheet
(Fixed to the floor
using bolts)
100
No. 2A201-519EN*A