Main Specifications For Site Planning - Toshiba Aquilion TSX-101A/I Series Site Planning Manual

Table of Contents

Advertisement

4.

MAIN SPECIFICATIONS FOR SITE PLANNING

(1)
Classification of the equipment
Class of protection against electric shock
Type of protection against electric shock
Level of protection against the ingress of
water
Type of safety when used in an environment
where flammable gas mixtures with air or
flammable anesthetic mixtures with oxygen
or nitrous oxide are present
Classification based on the operating mode
(2)
X-ray specifications
Pattern and type of X-rays
X-ray tube voltage
X-ray tube current
Scan time
*1: When high-power X-ray generator kit (option) is incorporated
(3)
X-ray tube leakage dose
X-ray tube leakage dose
X-ray beam shielding lead equivalent
X-ray tube inherent filtration
Inherent filtration other than the X-ray tube : Minimum 1.5 mm Al equivalent
(4)
Power supply requirements (for standard system)
See subsection 8.1 "Electrical Equipment".
(5)
Grounding
Grounding must be provided in accordance with all applicable legal requirements for medically
used electrical equipment.
(6)
Wiring work
Wiring work for the power supply must be provided as detailed in subsections 8.1 and 8.2.
Class I
Type B
Normal equipment (IPX0)
Equipment that should not be used in an
environment where flammable gas mixtures
with air or flammable anesthetic mixtures
with oxygen or nitrous oxide are present
Continuous operation with intermittent load
: Fan beam, continuous X-rays
: 80, 100, 120, 135 kV (±3%)
: 10 mA to 50 mA (±3 mA)
50 mA to 500 mA (±5%)
50 mA to 600 mA (±5%)
: (0.32), 0.5, 0.75, 1.0, 1.5, 2.0, 3.0 s
(0.32 s is for half scan)
: 1.0 mGy or less per hour
-5
(or 2.58 × 10
C/kg·H [100 mR] or less)
: 2.4 mmPb or more
: 1.0 mm Al equivalent or more
9
No. 2A201-519EN*A