Toshiba Aquilion TSX-101A/I Series Site Planning Manual page 44

Table of Contents

Advertisement

Power cables
Cable
Wiring
Connected
Connected
No.
pit
to: (1)
to: (2)
P01
P
System
Console
trans.
P02
P
System
Gantry
trans.
P04
P
Gantry
Console
CPU BOX
P00
P
Power
System
distribution
trans.
board
E01
P
Power box Gantry
P03
P
System
Gantry
trans.
E00
P
Power box System
trans.
Table 8-1 List of inter-unit cables (1/3)
Standard
Effective
Retract length
length (m)
length (m)
(m)
20
18
1.5 while
(10)
(8)
connecting to (1)
0.5 while
connecting to (2)
15
12
1.5 while
(11)
(8)
connecting to (1)
1.5 while
connecting to (2)
19
17
1.5 while
(14)
(12)
connecting to (1)
0.5 while
connecting to (2)
5
3
1.5 while
connecting to (1)
0.5 while
connecting to (2)
15
12
1.5 while
(11)
(8)
connecting to (1)
1.5 while
connecting to (2)
15
12
1.5 while
(11)
(8)
connecting to (1)
1.5 while
connecting to (2)
5
3
1.5 while
connecting to (1)
0.5 while
connecting to (2)
Number of cores ×
Maximum
Outer cable
2
length (m)
diameter (mm)
section area (mm
)
3 cores × 5.5
20
15
3 cores × 5.5
19
6 cores × 0.75
7
3 cores × 35
15
1 core × 35
3 cores × 35
15
7
1 core × 35
44
No. 2A201-519EN*A
Connector diameter (mm)
17.5
Solderless contact while
connecting to (1)
Solderless contact while
connecting to (2)
17.5
Solderless contact while
connecting to (1)
Open wire while connecting
to (2)
11.5
Solderless contact while
connecting to (1)
Connector (28 × 21 × 28)
plus solderless contact
35.2
Solderless contact while
connecting to (2)
12.2
Solderless contact while
connecting to (1)
Solderless contact while
connecting to (2)
35.2
Solderless contact while
connecting to (1)
Exposed cable at (2)
12.2
Solderless contact while
connecting to (2)
Remarks

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aquilion tsx-101a/h series

Table of Contents