Maintenance Space - Toshiba Aquilion TSX-101A/I Series Site Planning Manual

Table of Contents

Advertisement

5.4

Maintenance Space

(a)
Gantry
(b)
Patient couch
(c)
Console (in scanning room)
(d)
Console (in control room)
(1)
Since the scanner front cover opens outward, provide a clearance of 1400 mm for
opening/closing the cover.
(2)
Provide a clearance of at least 650 mm behind the scanner for servicing the LCSR, data
transfer section.
(3)
Provide a clearance of at least 800 mm for service.
(4)
Provide a clearance of at least 700 mm for servicing the servo amplifier etc.
(5)
Provide a clearance of at least 400 mm from the rear edge of the patient couch for
servicing the patient couch.
(6)
Provide a clearance of at least 500 mm with the CPU BOX pulled out toward the front
for servicing the console.
(7)
Provide a clearance of at least 500 mm with the REC BOX for servicing the REC BOX.
(8)
It is desirable to reserve a clearance of more than 400 mm for the use of the headrest.
Figure 5.4-1 Clearance required for maintenance
(2)
(3)
(a)
(b)
(1)
(c)
(7)
(5)
(d)
(6)
18
No. 2A201-519EN*A
(8)
(4)