Electrical Work; Electrical Equipment; Electrical Specification - Toshiba Aquilion TSX-101A/I Series Site Planning Manual

Table of Contents

Advertisement

8.

ELECTRICAL WORK

8.1

Electrical Equipment

8.1.1

Electrical specification

Application
Number of phases
Line voltage
Frequency
Power capacity
Rated capacity
(Actual load capacity)
Voltage variation due to
changing load
Line voltage variation
(*1) The cables between the power receiving transformer and the Aquilion power distribution
board must exceed the minimum nominal sectional area, obtained from the cable length
and the transformer capacity, as shown in figures 8.1-1 and 8.1-2.
In addition, independent cabling is desirable (the cable for the X-ray system (gantry
rotation section) should be separate from the cable for the gantry stationary
section/console).
(*2) This specification must be met even when a load is applied.
For X-ray system
(gantry rotation section)
*1
*2
Three phases
200 V
50 Hz or 60 Hz
100 kVA
85 kVA
5% or less
±10% or less
33
No. 2A201-519EN*A
Gantry stationary
section/console
14 kVA