Building Construction Work; Entrance Used To Bring-In The Unit - Toshiba Aquilion TSX-101A/I Series Site Planning Manual

Table of Contents

Advertisement

7.

BUILDING CONSTRUCTION WORK

7.1

Entrance Used to Bring-in the Unit

To carry the units into the room without any difficulty, conditions such as sufficient door size,
corridor width and ceiling height must be satisfied in advance.
(1)
Bringing-in method
The system is normally shipped with the covers mounted. If it is necessary to disassemble
the system to reduce the external dimensions or mass for any reason, consult Toshiba in
advance.
(2)
External dimensions and mass
The following figure shows the external dimensions and mass of the gantry and the required
size of the bringing-in entrance.
Height of the door
Width of the door
Load of the elevator
(3)
Carrying-in jig
To install the main body of the scanner easily, it is recommended that the carrying-in jig
should be used.
Recommended values
2050 mm or more
1100 mm or more
2 tons
Figure 7.1-1
21
No. 2A201-519EN*A
Mass: 1720 kg
Mass: 1750 kg