Signal cables (1/2)
Cable
Wiring
Connected
Connected
No.
pit
to: (1)
to: (2)
S01
S
Console
Gantry
REC BOX
S02
S
Console
Gantry
CPU BOX
(a) Gantry
S03
S
Console
(b) Speaker
CPU BOX
(Gantry)
(c) Speaker
(C-room)
S04
S
Console
Gantry
REC BOX
Table 8-1 List of inter-unit cables (2/3)
Standard
Effective
Retract length
length (m)
length (m)
(m)
19
16
0.5 while
connecting to (1)
1.5 while
connecting to (2)
19
16
0.5 while
(14)
(11)
connecting to (1)
1.5 while
connecting to (2)
19
16
0.5 while
(14)
(11)
connecting to (1)
19
16
2.0 while
(14)
(11)
connecting to (2)-
(a)
11
8
2.0 while
connecting to (2)-
(b)
2.0 while
connecting to (2)-
(c)
14
11
0.5 while
connecting to (1)
1.5 while
connecting to (2)
Number of cores ×
Maximum
Outer cable
2
length (m)
diameter (mm)
section area (mm
)
19
2-pairs, optical
fiber
19
18 pairs × 0.08
shielding
3 pairs × 0.18
19
shielding
2 pairs × 0.5
19
shielding
11
2 pairs × 0.5
shielding
19
2-pairs, optical
fiber
45
No. 2A201-519EN*A
Connector diameter
(mm)
Connector (22 × 8 × 25)
11
7.1
Connector (62 × 20 ×
52) while connecting to
(1)
Connector (40 × 12 ×
42) while connecting to
(2)
Connector (φ29 × 52)
4.2
while connecting to (1)
Connector (φ25 × 49)
5.1
while connecting to (2)-
(a)
5.1
11
Connector (22 × 8 × 25)
Remarks
Minimum
permissible
bending
radius:
110 mm
Minimum
permissible
bending
radius:
110 mm