Model 3336A/B/C
3-128. System Management Messages.
3-130. Remote. When it is first turned on, the -hp- 3336 is in the Local mode and under
front panel control. In order to be operated over the HP-IB, it must be switched to the
Remote mode. The Remote Message switches the instrument to the Remote mode. In this
mode, the only operational front panel controls are the Power switch and the Local key (see
Local Lockout Message, Paragraph 3-134). All other instrument functions are activated
over the HP-IB
through the system controller. The instrument configuration does not
change when switched to Remote.
3-133. Implementation.
The syntax and mnemonics for the program statement that im-
plements the Local Message are found in the controller documentation. Only the listen ad-
dress, which is $ for the -hp- 3336, is taken from the instrument documentation. An instru-
ment must be addressed to listen in order for it to enter the Local mode. A technical descrip-
tion of the Local Message implementation is presented in Figure 3-A-2 of Appendix A.
3-134. Local Lockout.
The Local Lockout Message disables the Local control on the front
panel of the -hp- 3336. All devices on the HP-IB with Local Lockout capability will respond
when this message is sent. The instrument can be switched to Local mode by executing the
Local Message. Local Lockout, however, will still be in effect if the instrument is switched
back to the Remote
mode.
To remove
Local
Lockout,
see Paragraph
3-136,
Clear
Lockout/Set Local Message.
3-135. Implementation.
The entire program statement that implements the Local Lockout
Message is found in the controller documentation.
No part of the program
statement
depends on the individual instrument. A technical description of the Local Lockout Message
implementation is presented in Figure 3-A-3 of Appendix A.
3-136. Clear Lockout/Set Local. This message switches all instruments on the HP-IB to the
Local mode and clears all Local Lockout conditions. Other methods to accomplish the same
thing are to disconnect the HP-IB cable, turn the controller off or turn the individual in-
struments off.
3-137. Implementation.
The
entire
program
statement
that
implements
the
Clear
Lockout/Set Local Message is found in the controller documentation. No part of the pro-
gram statement depends on the individual instrument. A technical description of the Clear
Lockout/Set Local Message is presented in Figure 3-A-4 of Appendix A.
Operation
- 3-23