Table Of Contents - HP 3336A Operating Manual

Synthesizer/level generator
Hide thumbs Also See for 3336A:
Table of Contents

Advertisement

Model 3336A/B/C
Performance Tests
4-8. PERFORMANCE TEST RECORD.
4-9. A Performance Test Record is located at the end of this section to help you consolidate
the test results, -hp- recommends that copies of the Performance Test Record be used.
Copies of the Performance Test Record can be made at any time without written permission
from Hewlett-Packard.
4-10. PERFORMANCE TESTS.
4-11. The following Performance Tests have been specifically developed to test the -hp-
3336:
Paragraph No.
Frequency Accuracy
4-12
4-14
4-16
Attenuator Accuracy
4-18
Phase Increment Accuracy
4-20
4-22
Output Balance
4-25
4-27
4-29
4-31
4-33
4-35
4-37
4-13. The frequency accuracy of the 3336 is not specified. However, the aging rate is. A
standard 3336 should pass this test, one year after the frequency has been calibrated. A 3336
with Option 004 should pass this test, one week after the frequency has been calibrated.
Specification:
(Aging Rate)
+5 x 10 -6 per year (20 to 30°C)
+5 x 10 -8 per week, Option 004
Electronic Counter
-hp- Model 5328A
a. Connect the equipment as shown in Figure 4-1.
4-11