HP 3336A Operating Manual page 107

Synthesizer/level generator
Hide thumbs Also See for 3336A:
Table of Contents

Advertisement

d. Set the electronic counter controls to measure the average time interval from input A
©
to input B (T.1.AVG AB). Set the counter to average 10' intervals. Note, the gate time to
average 107 intervals will be 1 second. Do not use the first reading displayed after changing
phase.
e. Adjust the 3336 (DUT) output phase until the counter displays a time interval from 190
ns to 210 ns. Each degree incremented on the 3336 will cause a 2.8 ns change in the counter
display.
f. Assign zero phase to the 3336 signal output. This is a shifted function of the PHASE
key.
i. Record the electronic counter reading with the test phase shift (T2).
"hp-3336(DUT)
Pour
ELECTRIC COUNTER
a44)
9)
545
eee
eee
752
OUT PUT
aa
TS Q
FEED THRU
TERMINATION
@
3336 -4-6