HP 3336A Operating Manual page 87

Synthesizer/level generator
Hide thumbs Also See for 3336A:
Table of Contents

Advertisement

Performance Tests
Table 4-1. Specifications (Cont'd)
Flatness: referenced to amplitudes at 10 kHz for the 50, 75 and 600 ohm outputs; referenced to amplitudes at 50 kHz
for the 124, 135 and 150 ohm outputs.
Model 3336A
10Hz
200
10K
50K
109K
2.09M
20.9MHz
+ .1dB(+
.07 dB with Option 005)
+ .12 dB
75 ohm output
150 ohm output
600 ohm output
Model 3336B
10Hz
200
10K.
50K
109K
2.09M
10.9M
20.9MHz
+ .1dB(+
.07 dB with Option 005)
+ .12 dB
75 ohm output
124 ohm output
135 ohm output
600 ohm output
-Model 3336C
10Hz
20.9 MHz
50 and 75 ohm outputs
+ .1dB(+
.07 dB with Option 005)
Attenuator Accuracy: (instruments without Option 005)
Attenuation
10Hz
1
M
20.9MHz
M
10
10 to 19.99 dB
+ .1dB
20 to 39.99 dB
+ .15¢dB
40 to 79.99 dB
+ .2dB
NOTE
Amplitude Accuracy is the sum of Absolute Accuracy and, as needed, Flatness and
Attenuator Accuracy.
Warm-Up Time:
30 minutes
MAIN SIGNAL OUTPUTS
On Carrier Return Loss:
Model 3336A
10Hz
10K
30K
2.09M
10.9M
2.09MHz
78 ohm output
150 ohm output
Model 3336B
10Hz
10K
30K
2.09M
10.9M
20.9MHz
75 ohm output
> 20 dB
Model 3336C
10Hz
TOK
30K
2.09M
10.9M
20.9MHz
*Return Loss of 50 ohm output is > 30 dB to 20.9 MHz with Option 005.
124 ohm output
135 ohm output
50 ohm output
75 ohm output
Model 3336A/B/C