Performance Tests
Model 3336A/B/C
d. Refer to the following table for Step e.
@ .
OUTPUT
AMPLITUDE FREQUENCY
NOMINAL
MINIMUM
MAXIMUM
3336A
750
—2.99 dBm
10 kHz
0.1941
0.1930
0.1952
150Q
—8.23 dBm
50 kHz
0.1502
0.1493
0.1510
6002
—2.99 dBm
10 kHz
0.5490
0.5459
0.5522
3336B
750
—2.99 dBm
10 kHz
0.1941
0.1930
0.1952
1240
—8.23 dBm
50 kHz
0.1365
0.1357
0.1373
1350
—8.23 dBm
50 kHz
0.1425
0.1416
0.1433
6000
—2.99 dBm
10 kHz
0.5490
0.5459
0.5522
3336C
502
—1.23 dBm
10 kHz
0.1941
0.1930
0.1952
75Q
—2.99 dBm
10 kHz
0.1941
0.1930
0.1952
e. Set up the 3336 for each of test conditions above and enter the ac voltmeter readings
on the Performance Test Record. Refer to the minimum and maximum values above to see
if the measured values meet the specifications.
4-16. Amplitude Flatness.
4-17. This performance test verifies that the -hp- Model 3336 meets the Amplitude Flatness
specification in Table 4-1.
Specification:
Referenced to amplitudes at 10 kHz for the 50, 75 and 600 ohm outputs. Referenced to
amplitudes at 50 kHz for the 124, 135 and 150 ohm outputs.
3336A
75 ohm output
+.1dB(+.07dB)*
10Hzto20.9MHz
150 ohm output
+.12 dB
10 kHz to 2.09 MHz
600 ohm output
+.25 dB
200 Hz to 109 kHz
3336B
75 ohm output
+.1dB(+.07dB)*
10Hzto 20.9 MHz
124 ohm output
+.15 dB
10 kHz to 50 kHz
+.1 dB
50 kHz to 10.9 MHz
135 ohm output
+.12 dB
10 kHz to 2.09 MHz
600 ohm output
+ .25 dB
200 Hz to 109 kHz
3336C
50 ohm output
+.1dB(+.07dB)*
10Hzto 20.9 MHz
75 ohm output
+.1dB(+.07dB)*
10Hzto 20.9 MHz
*NOTE
Specifications in parenthesis apply to instruments with Option
005.
Required Equipment:
75 ohm, .5 V Thermal Converter
DC Voltmeter
-hp- Model 11051A/H07
-hp- Model 3455A
4-14