Model 3336A/B/C
Performance Tests
NOTE
When making harmonic distortion measurements with a spec-
trum analyzer, make sure that the analyzer's harmonic distor-
tion does not mask the distortion of the device under test. One
technique to ensure this is to increase the analyzer's input at-
tenuation, which results in lower signal levels at the analyzer's
input. This yields better intermodulation and harmonic distor-
tion performance. Adjust the analyzer's reference level controls
to obtain the proper display. For good harmonic distortion
performance in the -hp- Model 141T/8552B/8553B,
the signal
level at its mixer must be less than
— 40 dBm. If you are using
this analyzer, set its input attenuation to 40dB.
a. This test will require two different setups: one for frequencies equal to or less than 10
kHz and one for frequencies greater than 10 kHz.
b. Set the 3336 controls as follows:
OUTPUT
sc iiuiaiern
whack haw eeuss 75 ohm
AMPLITUDE es ocsneoriccet a 5.00 dBm
FAST LEVELING
2. cece ncseereieess OFF
c. Perform Steps d through f for each configuration listed below.
3336 FREQUENCY | FAST LEVELING
SETUP
10 Hz
OFF
30 Hz
OFF
50 Hz
OFF
Figure 4-9A
10 kHz
OFF
10 kHz
ON
100 kHz
OFF
100 kHz
ON
1 MHz
OFF
1 MHz
ON
Figure 4-9B
5 MHz
OFF
5 MHz
ON
20.9 MHz
OFF
20.9 MHz
ON
d. Tune the spectrum analyzer to display the fundamental frequency and at least four of
its harmonics.
e. Adjust the analyzer's input sensitivity controls until the amplitude of the fundamental
is 0 dB (full scale display).