HP 3336A Operating Manual page 16

Synthesizer/level generator
Hide thumbs Also See for 3336A:
Table of Contents

Advertisement

Make sure the correct fuse is installed for the selected line voltage!
Line Voltage
100/120 Vac
220/240 Vac
-hp- Part Number
1 Amp
2110-0001
Y% Amp
|
2110-0012
USE FAST BLOW TYPE FUSES ONLY! Using slow blow type
Juses or fuse values other than those recommended defeats an im-
portant protection circuit and will damage the Model 3336.
2-10. Power Cable.
2-11. In accordance with international safety standards, this instrument is equipped with a
three-wire cable. When connected to an appropriate power line outlet, this cable grounds the
instrument cabinet. The type of power cable shipped with each instrument depends on the
country of destination. Refer to Figure 2-1 for the connector configuration and -hp- part
number of the available power cables.
280 V
OPERATION
PLUG*: SEV 1011.1959-24507
PLUG*: CEE?-V1t
PLUG*: CEE22-V1
TYPE
CABLE*: HP 8120-1692
CABLE*: HP 6120-1860
CABLE': HP 8120-2856
CABLE*:
HP 8120-2104
125 V-6a"°
128 V-6Aa°°
"&
PLUG*: NEMA 1-16P
PLUG*:
NZSS 19B/AS C112
PLUG*: BS 1363A
PLUG*:
NEMA 5-15
CABLE*: HP 8120-0684
CABLE*: HP 8120-0698
CABLE*: HP 8120-1703
CABLE*: HP 8120-0698
. CABLE*: HP 8120-1621
* The number shown for the plug ie the industry identifier for the plug only.
The number shown for the cable is en HP pert number for e complete cable including the plug.
* UL Hetad for use in the United States of America
Figure 2:1. Power Cables.
2:12. HP-IB Connections.
2-13. Interconnection data concerning the rear panel HP-IB connector is provided in Figure
2-2. This connector is compatible with the -hp- 10833 (A,B,C, or D) HP-IB cables. The lengths
of these cables are as follows:
10833A
1 meter
10833B
2 meters
10833C
4 meters
10833D
.5 meters