GE H
DRAFT
EALTHCARE
D
KTD102576, R
IRECTION
4-5-2-2
Load "Complete Backup"
NOTE:
Following procedure should be used, if the user prefers factory default settings, which are adapted for
the installed Application Software version.
1.) Perform
2.) Select the "Complete Backup" (marked blue; see:
copy the Complete Backup into the Load Data field.
Complete Backup
3.) Click the LOAD button to start the loading procedure of the complete backup into the system.
The US Application Software restarts.
7
DRAFT (A
EVISION
Preparations
on page 4-34.
Figure 4-25 Load Backup Data
Chapter 4 - Functional Checks
23, 2012)
UGUST
Figure 4-25
V
E8 / V
OLUSON
OLUSON
S
ERVICE
below) and click the >> button to
>> button
[Load] button