GE Voluson E8 Service Manual page 309

Hide thumbs Also See for Voluson E8:
Table of Contents

Advertisement

GE H
DRAFT
EALTHCARE
D
KTD102576, R
IRECTION
File Decompress Error. Action: Load volume files from other storage medium.
File Decompress method Unknown.
Action: Load volume files from other storage medium.
File End Error. Action: Load volume files from other storage medium.
File Memory Missing. Action: Load volume files from other storage medium.
File Not Found. Action: Load volume files from other storage medium.
File Pos. Action: Load volume files from other storage medium.
File Read Error. Action: Load volume files from other storage medium.
File Type Unknown. Action: Load volume files from other storage medium.
File Volume size not consistent.
Action: Load volume files from other storage medium.
FilterBuffer " Memory allocation failed". Action: please restart the system
GeckoBF: Wave Form Table wasn't found! Unknown Wave Form Table Key!
Action: Restart the System!
GeoDescription3D_TissueCF::operator=: handed over argument is not of same
type, dynamic cast failed!!
GeoDescriptionMotion Constructor: unknown mode Parameter handed over
GeoDescriptionMotion::operator=: handed over argument is not of same type,
dynamic cast failed!!
GeoDescriptionPWMode::operator=: handed over argument is not of same type,
dynamic cast failed!!
GeoDescriptionTissueCF Constructor: unknown mode Parameter handed over
GeoDescriptionTissueCF::operator=: handed over argument is not of same type,
dynamic cast failed!!
Hardware doesn't support CW-mode
hardware error on <var> Action: restart the system
HardwareRelatedSoftware_Windows in write have different ProbeAcousticUnitIDs.
Action: disconnect all connected probes and connect again,
if not ok restart the system
HardwareRelatedSoftware_Windows in write have different ProbeScanFuncIDs.
Action: disconnect all connected probes and connect again,
if not ok restart the system
IBegrenzer.cpp BCshots no TxPulseEntrys.
Action: restart the system, If after restart not ok then call technical support
IBegrenzer.cpp Bshots no TxPulseEntrys.
Action: restart the system, If after restart not ok then call technical support
IBegrenzer.cpp Bshots TxMultiFocus problem
7
DRAFT (A
EVISION
Error Messages
Chapter 7 - Diagnostics/Troubleshooting
23, 2012)
UGUST
load volume files from other storage medium
load volume files from other storage medium
load volume files from other storage medium
load volume files from other storage medium
load volume files from other storage medium
load volume files from other storage medium
load volume files from other storage medium
load volume files from other storage medium
load volume files from other storage medium
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
restart the system
pencil probe + CW-Hardware not available->
restart the system
restart the system
disconnect all connected probes and connect them again;
if not ok restart the system
disconnect all connected probes and connect them again;
if not ok restart the system
restart the system
restart the system
restart the system
V
E8 / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
Actions
7-37