GE H
DRAFT
EALTHCARE
D
KTD102576, R
IRECTION
5-15-3-2
Update
5-15-3-2-1
FMI from DVD
By means of the FMI FROM DVD button, the Systems C:\ image is partly or completely updated.
The System Software parts to be upgraded depend on contents of the used System DVD.
PRECONDITION:
The first "Boot Device" in BIOS has to be Hard Disk Drive.
NOTE:
During "FMI from DVD" the used system configuration (incl. Full Backup) will be stored on R:\.
If required, the previously used System configuration (before FMI from DVD was performed) can
be restored by activating the "Rollback" function. Operation see:
5-15-3-2-2
ASI - Additional Software Installation
Click the ASI button to install additional software (e.g., Process Logger).
The Software parts to be installed depend on the contents of the System DVD that is used.
5-15-3-2-3
Restore Software
Click the RESTORE SOFTWARE button to restore the system software from your hard disk.
5-15-3-3
TCP/IP Buffersize
The TCP/IP Buffersize selects the amount of buffer memory used for DICOM transfers (both directions).
5-15-3-4
Activate / Deactivate Service...
This offers GE technicians the possibility to view the entire customer´s desktop and operation system.
Using VCO (Virtual Console Observation) a service technician or the OnLine Center can access and
modify all PC settings and programs or run diagnostics on the customer´s ultrasound scanner.
Operation see:
5-15-3-5
Common Service Desktop (CSD)
Access to the Common Service Desktop (CSD) by entering security user level and password.
Activate the service platform as described in
5-15-3-6
Request for Service
Click the RFS button, fill out the "Request For Service" form and then send the problem description to
GE Service/Application representatives.
Operation see:
5-15-3-7
Delete all Patients
1.) Click the DELETE ALL PATIENTS... button. Following WARNING message appears on the screen.
WARNING
WARNING
If you select the YES button, all patients data, studies, images and measure report data
will be deleted permanently from the hard disk and cannot be recovered!
!! WARNING:
7
DRAFT (A
EVISION
Section 7-5 "Remote Access to Service Platform" on page 7-12
Section 7-2-2 "Request for Service (RFS)" on page 7-5
Figure 5-50 Warning message
Chapter 5 - Components and Functions (Theory)