6-3-3
Color Calibration
1.) Press the UTILITIES key on the control panel.
2.) In the "Utilities" menu touch SYSTEM SETUP to invoke the setup desktop on the screen.
3.) On the left side of the screen select ADMINISTRATION and then click the
4.) Enter the password
5.) Select the MONITOR TEST button in the "Service Tools" menu.
6.) Confirm the "Monitor Test" message with YES.
The Monitor Test main menu appears on the screen.
7.) Select the "Color calibration" icon and press the
-
8.) Press the
-
9.) To return to the Main menu, press the
10.)To exit the Monitor Test program, press EXIT.
7
DRAFT (A
EVISION
SHE
and click the ACCEPT button.
Figure 6-5 Monitor Test - Main menu
WHITE is displayed without any tint (discolor) or colored pixels.
right/left trackball
key repeatedly to step through RED, GREEN and BLUE.
Each color is displayed correctly (without any tint or discolored pixels).
Chapter 6 - Service Adjustments
23, 2012)
UGUST
right/left trackball
upper trackball
key.
V
E8 / V
OLUSON
OLUSON
S
ERVICE
SERVICE
tab.
color bars
color bars
raster lines
gray scale
Color calibration
icon
EXIT
(screen becomes white).
E6
M
ANUAL
6-5