Software and Functional Checks after Installation/Upgrade Procedure
1.) Press the UTILITIES key on the control panel.
2.) In the "Utilities" menu touch SYSTEM SETUP to invoke the setup desktop on the screen.
3.) On the left side of the screen select ADMINISTRATION and then click the
4.) Verify the correct settings of the
If necessary, customize the settings according to the printout.
5.) Click the
Voluson E8
Application Software
version
Backup File Name
Tune version
of Application presets
start Hardware
configuration detection
Figure 8-14 System Setup - Administration - SYSTEM INFO page (e.g., BT13)
6.) Check the Application Software version.
7.) Check that the Tune version of the Application presets match the Application Software version.
It is neither required nor advisable to reload a previously stored "Full Backup" after a software
NOTICE
upgrade that was performed by means of the FMI FROM DVD button!
!! NOTICE:
If the Tune version does not match the Application Software version, a warning message appears
whenever booting up the system.
In this case, it is essential to load the proper Application Settings (image presets), adapted for the
installed software version. Refer to:
8.) Disconnect all probes and then click the button at the bottom of the
Hardware configuration detection (see:
9.) Perform basic functional checks to ensure system is functioning normally.
7
DRAFT (A
EVISION
SYSTEM INFO
tab.
Chapter 8 - Replacement Procedures
23, 2012)
UGUST
page; see:
Figure 8-2 on page
OPTIONS
Section 5-15-3-11 "Load Application Settings" on page
Figure 8-14
above).
V
E8 / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
OPTION
tab.
8-5.
5-74.
page to start
SYSTEM INFO
E6
8-11