GE Voluson E8 Service Manual page 244

Hide thumbs Also See for Voluson E8:
Table of Contents

Advertisement

GE H
RAFT
EALTHCARE
D
KTD102576, R
IRECTION
Section 5-10
Power Distribution
5-10-1
RTN - Primary Power Module (AC/AC)
C20
(16A type)
C14
(10A type)
5-10-1-1
Mechanical Concept and Overview
The AC Power's main tasks are to supply the various internal subsystems with AC power and to
galvanically isolate the scanner from the on site Mains Power System. To reduce inrush current, an
inrush current limiter is implemented.
Voltage to peripherals can be configured to either 115 VAC or 230 VAC.
From the Mains Power Input module, the AC power is routed via an Inrush Current Limiter to an internal
outlet connector for the Mains Transformer.
Connector C14 (10A type) is installed by default.
NOTICE
Note: If the Mains Voltage is lower than 200V, Connector C20 has to be installed!
!! NOTICE:
5-10-1-2
Major Functions of RTN
Inrush Current limiter
The RTN Primary Power Supply module generates 34.5VAC (+/-2V) as an input voltage for the
Secondary Power Supply module (RTP).
The RTN module contains also the isolation transformer for the peripherals (AUX).
(Maximum load: 345VA) see: Basic User Manual, Technical Data/Information
NOTE:
All DC-supply voltages for built in peripherals are generated in RTP-module not inside RTN.
NOTICE
The systems mains supply input voltage can be set to: 100V, 115V, 125V, 225V and 250V.
Modifications of voltage settings has to be done by an authorized service person only!
!! NOTICE:
The output voltages may be set to 115VAC or 230VAC (independent from the input voltages).
NOTE:
The "Mains voltage indicator" on the RTN primary power supply should be set in accordance to the input
voltage level. This indicator is NOT a voltage selection switch!
5-60
7
DRAFT (A
EVISION
Lift System (230V)
Figure 5-34 Primary Power Module - RTN
Section 5-10 - Power Distribution
23, 2012)
UGUST
Auxiliary Output (115V/230V)
V
E8 / V
OLUSON
OLUSON
S
M
ERVICE
ANUAL
1 screw
Mains voltage indicator
(shows input voltage level)
Circuit breaker
Protective Earth connection
Equipotential connection
pull-out protection
Mains Power cable
E6

Advertisement

Table of Contents
loading