GE Voluson E8 Service Manual page 72

Hide thumbs Also See for Voluson E8:
Table of Contents

Advertisement

GE H
RAFT
EALTHCARE
D
KTD102576, R
IRECTION
Table 3-4
Step
3.
Remove the left and right side walls, finally remove also the rear panel of the crate; the rear panel is used for the ramp.
4.
Carefully remove foam packing material and plastic bag from the ultrasound system and monitor.
Caution: Two people are needed in the next step due to the weight of the equipment.
5.
Disengage the brakes, grasp the system at the handle, then slowly move system down the ramp (bottom side of the rear panel).
rear panel
(bottom side up)
is used for ramp
Note: Packing crate and material should be stored for future use.
3-6
7
DRAFT (A
EVISION
Unpacking Procedure
remove side walls
Section 3-3 - Receiving and Unpacking the Equipment
23, 2012)
UGUST
Task
V
E8 / V
OLUSON
S
ERVICE
finally remove rear panel
(used for the ramp)
NOTICE
!! NOTICE:
T
L
HE SYSTEM IS
OWERED TO ITS
,
MINIMUM HEIGHT
WITH MONITOR
L
FLAPPED AND
OCKED DOWN
T
C
C
HE
ONTROL
ONSOLE IS
CENTERED AND LOCKED IN POSITION
E6
OLUSON
M
ANUAL
.
.