5.1 Checking the Performance and Specifications

5.1.2 Preparation of Operation

(3)
Performance Check Parts Attached to ChromasterUltra Rs
6270 Autosampler
 Sample vial, cap, septum
 Thermo rack
NOTE:
Performance specification should be checked in the
following combination.
Part No.
080-3595
855-2351
080-3704
In the other cases (refer to Section 7.3), adjust the rack
parameters according to the combination used for your
practical application (refer to the maintenance software
instruction manual), and check the reproducibility
performance before proceeding to operation.
NOTE:
For good reproducibility, avoid using a septum repeatedly
to prevent a cause of a hole. It causes a fluctuation of
sample concentration in sample vial.
NOTE:
Be sure to use the sample vials, caps and septa
recommended by Hitachi High-Tech Science (refer to
Section 7.3) If other parts are used, the instrument may
be malfunction or its original performance may not be
available.
(1)
Figure 5-1 shows how to connect the units for performance
check. According to this figure, carry out tubing.
Before energizing the system, ensure that the tubing between
the autosampler and pump/detector is properly connected.
5 - 2
Part Name
1.5 mL Sample vial
Septum (Si -TEF), Slit (-)
Screw cap