5.3 Consumables and Spare Parts
Figure 5-12 Connecting the Plastic Syringe (or Tube)
(2)
Connect the provided plastic syringe to the port4.
Plastic syringe
(3)
Start maintenance software. Click Maintenance -
Autosampler tab.
(4)
Loosen the plunger setscrews with your hand to detach the
syringe from the hook.
(5)
Click Syringe Change.
(6)
Turn the top end (threaded metal part) of the syringe
counterclockwise with your hand to separate the syringe from
the syringe mount. Pull the plunger frontward out of the hook.
(7)
Before mounting a new syringe, put distilled water (or wash
fluid) into it. Avoid air bubbles in the plunger.
Plunger groove
Plunger
Syringe
Beaker
Figure 5-13 Filling with Distilled Water
(8)
Mount the syringe (secure the top screw after engaging the
plunger groove with the hook). Be sure to engage the plunger
groove with the hook positively.
5 - 28
Fill the syringe with distilled
water by pulling the plunger.
Distilled water
(or wash fluid)