8.2.4 Dual Channel Level

This template adds Level functionality to the Dual Channel PV
template.
For each channel, this consists of applying a Specific Gravity
calculation to the PV to provide the level of a liquid within a
vessel.
A volume calculation can also be applied to the level to
provide the volume contained within a vessel.
84
1
/
DIN
8
8 Templates and Functionality
Specific
Gravity
AIN 1
Calculation
Specific
AIN 2
Gravity
Calculation
L
a
b
Cube:
Volume = a x b x L
Volume Constant = a x b
PV 1
(Level)
Volume
Volume
Calculation
PV 2
(Level)
Volume
Volume
Calculation
L
d
Cylinder:
Volume =  x (
d
2
/2)
x L
Volume Constant =  x (
d
2
/2)