Integral Transmitter - Probe Cable Connections - ABB Endura AZ series User Manual

Integral and remote transmitter, combustion oxygen monitor
Table of Contents

Advertisement

Endura AZ series integral and remote transmitter
Combustion oxygen monitor
4.10 Integral Transmitter – Probe Cable Connections
Terminal / Cable Color
Blue
Brown
Screen
White
Green
Black
Red
Grey
Violet
White / Blue
White / Red
White / Green
White / Orange
White / Black
White / Yellow
Table 4.6 Probe Cable Connections at Integral Transmitter
Standard Connections
H (Blue)
H (Brown)
SCN (Screen)
TC – (White )
TC + (Green)
Cell – (Black)
Cell + (Red)
ACJC (Grey)
ACJC (Violet)
Fig. 4.12 Probe Cable Connections at Integral Transmitter
IM/AZ20E–EN Rev. F
Tag ID
Connection Type
H
Heater
H
Heater
SCN
Screen
TC –
Thermocouple (–ve)
TC+
Thermocouple (+ve)
CELL –
Oxygen Input (–ve)
CELL +
Oxygen Input (+ve)
ACJC
PT1000 Cold Junction Compensation
ACJC
PT1000 Cold Junction Compensation
SV2
Solenoid Valve – Gas 2
SV COM
Solenoid Valve – Common
SV1
Solenoid Valve – Gas 1
PS1
Pressure Switch – Gas 1
P COM
Pressure Switch – Common
PS2
Pressure Switch – Gas 2
4 Electrical Installation
*AutoCal Connections
SV2 (White / Blue)
SV COM (White / Red)
SV1 (White / Green)
PS1 (White / Orange)
PS COM (White / Black)
PS2 (White / Yellow)
* For non-AutoCal
versions the 6-way
terminal block is not fitted
23

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Endura az25Endura az20Endura az10

Table of Contents