ABB Endura AZ series User Manual page 49

Integral and remote transmitter, combustion oxygen monitor
Table of Contents

Advertisement

Endura AZ series integral and remote transmitter
Combustion oxygen monitor
Parameter
...Device Setup
Security Setup
Standard
Advanced
Reset Fact. Settings
Reset Temp. Trip
Table 5.9 Device Setup Menu (Continued)
IM/AZ20E–EN Rev. F
Comment / [Range]
Used to set Standard and Advanced level passwords of
up to 6 alphanumeric characters.
Note. Standard and Advanced passwords are not set at
the factory and must be added by the end-user(s).
Set by Standard and Advanced users.
Set by Advanced users.
Resets all configuration parameters to their default values.
For heated probes (AZ20 and AZ30), if the cell
temperature exceeds a preset limit (approx. 850 °C
[1562 °F]), a hardware over- temperature trip automatically
activates and switches off power to the heater.
If the cell temperature remains at this elevated
temperature for less than 30 minutes, the trip is reset
automatically.
If the elevated temperature remains for more than 30
minutes, it is necessary to reset the trip (at this menu).
Alternatively, the trip can be reset by switching power to
the transmitter off and on again.
5 Programming
Default
None
Factory-set
N / A
47