Endura AZ series integral and remote transmitter
Combustion oxygen monitor
Parameter
...Easy Setup
Cable Length
Autocal Hardware
None
Internal
Remote
%O2 Range Hi
%O2 Range Lo
Thermocouple Type
K
B
N
R
S
Table 5.6 Easy Setup Menu (Continued)
34
Comment / [Range]
The length in metres of the cable between the probe and
the remote transmitter. This length is used in the cold
junction measurement to compensate for the impedance of
the cable.
[0 to 100 m]
Note. Applicable only to remote transmitters.
Selects the type of automatic calibration hardware to be
used.
All auto-calibration functions disabled.
Select this option if the probe's (optional) built-in automatic
calibration is fitted.
Select this option if an external automatic calibration system
is to be used.
Note. The solenoid valve outputs or relays can be used to
activate a remote automatic calibration system.
Sets the maximum oxygen concentration.
[0.01 to 100 % O
]
2
Sets the minimum oxygen concentration.
[0.01 to 100 % O
]
2
Selects the thermocouple type used for cell temperature
measurement.
For heated probes (AZ20 and AZ30) this is fixed as Type K.
For unheated probes (AZ25 and AZ35), select the
thermocouple type used in the probe being used with the
transmitter.
[–100 to 1300 °C (–148 to 2372 °F)]
[–18 to 1800 °C (–0.4 to 3272 °F)]
[–200 to 1300 °C (–328 to 2372 °F)]
[–18 to 1700 °C (–0.4 to 3092 °F)]
[–18 to 1700 °C (–0.4 to 3092 °F)]
5 Programming
Default
0 m
None
25.00 %O
2
0.01 %O
2
K
IM/AZ20E–EN Rev. F