9.7 Remote HMI (Human Machine Interface)

To run the Remote HMI application:
1. Launch the Remote HMI application:
2. From the Options / Comms Ports menu, select the COM number assigned in the Service Port
Switch.
3. Click Save Settings.
4. Operate the Remote HMI using the mouse and cursor to select the
the same way as operating the physical HMI.

9.8 HART Client

Note. Any standard HART application can be used via the infrared interface and the ABB Service Port
Switch. All communications are handled automatically with no need for a HART modem.
To run the HART client:
1. Configure the HART application to communicate with the COM port specified previously in the
Service Port Switch – see Section 9, page 91.
2. Launch any preferred PC HART application.
3. Configure the COM port on the running HART application to be the COM port specified previously in
the Service Port Switch.
The HART application operates as normal allowing any HART command to be read or written.
Note. An increase in operating speed may be noticed as HART communication transmits faster through
the service port than through a normal HART modem.