Endura AZ series integral and remote transmitter
Combustion oxygen monitor
...Calibration Procedures
<Cal Type>
O2
1.00 %O2
Cal Offset
0.00 mV
Cal Factor
1.000
Reject
• New calibration
• New calibration
coefficients
discarded
Save successful
•
Accuracy
•
check entry added
to Performance
Log
• Return to Sensor
• Return to Sensor
Cal.
Save unsuccessful
•
• Display Failure
<Cal Type>
O2
20.95 %O2
Calibration Failed
Signal Not Stable
Exit
• Return to Sensor
Cal.
IM/AZ20E–EN Rev. F
Calibration completed successfully
Accept
coefficients saved
Calibration
entry added to
Performance Log
Cal.
Accuracy
check entry added to
Performance Log
Calibration failed to complete
The reason for failure is displayed:
Signal Not Stable
not met
No Response
of a test gas detected
NV Error
new calibration factor unsuccessful
Result Out of Bounds
calculated calibration factor is outside the limits defined
by Factor Limit and / or the newly calculated calibration
offset is outside the Offset Limit parameter – see page
42
– no response to the application
– calibration successful but Save of
.
6 Calibration
– stability criteria
– newly
67