Endura AZ series integral and remote transmitter
Combustion oxygen monitor
Parameter
...Easy Setup
Temperature Units
°C
°F
Factory Cal. Offset
Factory Cal. Factor
New Probe/Cell Fitted Used to confirm and setup a new probe or cell.
Table 5.6 Easy Setup Menu (Continued)
IM/AZ20E–EN Rev. F
Comment / [Range]
Selects the units to be used for all temperature values in the
transmitter.
Used to enter the calibration offset value supplied with the
new probe / cell.
When changing the Probe Type to / from AZ10, the
following defaults are used:.
AZ10 = –10 mV
non-AZ10 = 0 mV
Note. Refer to the label supplied with cell.
Used to enter the calibration factor value supplied with the
new probe / cell.
Note. Refer to the label supplied with cell.
Select OK to confirm a new probe or cell has been fitted;
this resets the diagnostic measurements and states.
A new entry is created in the Performance Log to record
that a new cell or probe has been fitted and that factory
values have been entered.
5 Programming
Default
°C
1.00
N / A
35