ABB Endura AZ series User Manual page 31

Integral and remote transmitter, combustion oxygen monitor
Table of Contents

Advertisement

Endura AZ series integral and remote transmitter
Combustion oxygen monitor
Display Overview
Signals View
O2
O2
xx.x %O2
xx.x %O2
TMP
xxx °C
dT
x.x deg C/m
Back
Signals Units
Mnemonic
Displayed
O2
%O2
TMP
°C or °F
dT
°C / Min or °F / Min
mV
mV
O/P
%
CJ
°C or °F
Off
mV
Fct
zOf
mV
Dev
%O2
K
d
K / Year
RT
s
RR
%O2 / s
Vac
V
Frq
Hz
Int
°C or °F
Table 5.3 Operator Menu Screens (Continued)
IM/AZ20E–EN Rev. F
Section and Function
Signals View
The Signals View displays active signals and their values as a list.
Values for signals that have not been configured are not
displayed.
Exit
Use the
Description
% Oxygen
Cell Temperature
Rate of Rise of
Cell Temperature
Cell Millivolts
Heater Output
Cold Junction Temp
Calibration Offset
Calibration Factor
Cell Zero Offset
O
Deviation
2
Cell Impedance
Cell Impedance Drift
Test Gas Response Time
Recovery Rate
Mains Voltage
Mains Frequency
Internal Temperature
and
keys to scroll through active signals.
Typical Values
0.01 to 100.00 % O
>700 °C (1292 °F)
After stabilization phase
0 to 60 °C / min
(0 to 108 °F / min)
–50 to 190 mV
0 to 100 %
–20 to 70 °C (–4 to 158 °F)
–20 mV to 20 mV
0.900 to 1.100
<20 mV
<0.5 % O
0.1 to 10.0 k
<1 k
8 to 30 s
0.05 to 0.30 % O
90 to 265 V
50 to 60 Hz
–20 to 70 °C (–4 to 158 °F)
5 Programming
2
2
2
29