4.4.2 Two-Point Scaling

The span (inclination) and offset of the displayed value can be defined arbi-
trarily.
Set P1 (0.5mm) at "0.0000" and P2 (2mm) at "3.0000".
Display
2.0000
0
・Enter the first and second points.
・Enter the detected value on the upper line and the value to be displayed
on the lower line.
If the setting is incorrect, an error is displayed. Press the
cancel the error and repeat entry.
<Setting error>
・The detected value is beyond the detection distance range.
・The detection value of the first point is equal to or larger than the
POINT
detection value of the second point.
・The displayed value of the first point is equal to the displayed value of the
second point.
・The span (inclination) becomes more than the value of (16 devided by the
detection range) times.
(Scaling is possible up to 16 times for 1mm type, and 3.2 times for 5mm
type.)
・To reverse the span (inclination), select "SLOPE" of one-point scaling or
use the INV function.
78
Display
3.0000
2.0000
P1
0
2mm
P2
0.5mm
2mm