Two-Point Scaling - Panasonic GP-X Series User Manual

Digital displacement sensor
Hide thumbs Also See for GP-X Series:
Table of Contents

Advertisement

4.5.2 Two-Point Scaling

The span (inclination) and offset of analog voltage output can be set arbi-
trarily.
Set P1 (0.5mm) at "0.0000" and P2 (2mm) at "3.0000".
Voltage
(V)
5.0000
0
・Enter the first and second points.
・Enter the detected value on the upper line and the displayed value on the
lower line.
If the setting is incorrect, an error is displayed. Press the
cancel the error and repeat entry.
<Setting error>
・The detected value is beyond the detection distance range.
・The detection value of the first point is equal to or more than the detection
POINT
value of the second point.
・The voltage output value of the first point is equal to the voltage output
value of the second point.
・The span (inclination) becomes 3.2 times the default inclination or more.
・To reverse the span (inclination), select SLOPE" of one-point scaling or
use the INV function.
Voltage
(V)
5.0000
3.0000
P1
0
2mm
-1.0000
P2
0.5mm
2mm
key to
83